Steinfurth Offline Carbo Analyzer (OLCA)
Automatic CO2 monitoring in beverage lines, tanks and kegs
 |
|
|
Inline/Offline Carbo Analyzer (ILCA/OLCA)
|
- Inline/Offline Carbo Analyzer (ILCA/OLCA) - The OLCA works based on the manometric method and measures the CO2 concentration in tanks or beverage kegs.
The tank or keg is connected to the OLCA, and the beverage sample is transferred to the measuring block. The device does the sample extraction automatically.
A series of samples are possible and for each sample a small amount of liquid (ca. 200-300 ml) is sufficient. Each sample is degasified in the measuring chamber. Using beverage specific formula, OLCA calculates the CO2 concentration from pressure and temperature in equilibrium. The CO2 value is shown on the LCD. Optionally, CO2 concentration, pressure and temperature can be transferred to the PC via RS232 or RS 485 serial interface. |
|
|
Details |
|
|
| |
| |
| |
| |
|
|
Technical Information |
Advantages |
|
Measuring range: 2..12g CO2/l (1..6 vol) Reproducibility: 0,07 g/l Parameter sets: for up to 8 different products Output: every 45 sec, independent from previous result, by LCD Serial interface: RS 232 or RS 485 Signal output: current signal 4..20mA, voltage signal 0..10V, ProfiBus (1.2002) Supply voltage: 120 V / 230 V AC, 50 Hz Protection class: IP 65 |
- Manometric method
- Automatic sampling
- Programmable measuring cycle
- CIP resistant
- Easy maintenance
- No pumps etc. necessary
|
|
|